Revision C 250cx Series Maternal/Fetal Monitor 16-5
2036946-001
Troubleshooting: FECG Troubleshooting
FECG Troubleshooting
External Uterine Activity Troubleshooting
FECG Troubleshooting
Problem Probable Cause Possible Solution
Internal FECG erratic or not recording
properly.
Cable not properly connected to monitor.
Attachment pad or legplate not securely
attached to patient.
Electrode wire not secure in
legplate post.
Paste is dried or incorrect paste is being
used.
Electrode not properly attached.
No FECG signal.
Defective electrode.
Defective attachment pad.
Service required.
Ensure cable is securely attached to the
monitor.
Secure attachment pad or legplate to
patient.
Inspect legplate connection.
Check ECG paste; re-apply, if
necessary.
Replace electrode.
Auscultate FHR.
Replace electrode.
Replace attachment pad.
Call Biomedical Engineering
Department.
Rate in FHR area of the display and the
FHR trend on the strip chart paper do not
correlate.
Monitor is set for 30 bpm/cm vertical scale
and 20 bpm/cm vertical scale strip chart
recorder paper is being used (or vice
versa).
Call Biomedical Engineering Department.
External UA Troubleshooting
Problem Probable Cause Possible Solution
Tocotransducer not recording contractions.
Transducer not properly connected to
monitor.
Transducer not properly placed.
Transducer not secured to patient.
Defective transducer/cable assembly.
No maternal contractions.
UA Reference range exceeded.
Ensure that transducer is securely
attached to monitor.
Reposition transducer.
Secure or re-apply transducer to patient.
Replace transducer/cable assembly.
Wait.
Loosen belts or remove transducer from
patient. Press
UA Reference button while
no pressure is applied to transducer
button. Re-apply transducer. Do not
overtighten belt. Press
UA Reference
button again between contractions. (See
“Out of Range Condition” on page 6-4 for
further information.)