Revision C 250cx Series Maternal/Fetal Monitor C-9
2036946-001
Spectra Alerts: Alert Parameters Summary
Alert Parameters Summary
The FHR setup screens (FECG or US/US2) include an Alert Parameters summary
presenting an overview of the alert analysis results for FHR and UA. Fields shown
in inverse video are contributing factors to an alert. Some fields contain a second
column to provide qualifying information. The Table, “Summary of Alert
Parameters”, provides a list of possible results which can appear in the Alert
Parameters summary.
NOTEParameters displayed in inverse video are contributing to an active alert.
Items displayed in normal video may indicate an alert is pending.
Summary of Alert Parameters
Parameter Label
Alert Analysis Result Possibilities
Column 1 Column 2
Alert Level
Baseline HR (bpm)
Range
(For example: 145–150)
High
Low
FHR Variability Unknown
Average
Increased
Decreased
Flat
Decelerations Absent
Present
Artifact/Arrhythmia Present
UA Baseline Pressure
pressure in relative units or mmHg
(kPa when selected)
Hypertonus
UC in 10 min
# of UCs
Tachysyst
Brady
Tachy
Increased
Decreased
Flat
Present
Hypertonus
Tachysyst

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD