Revision C 250cx Series Maternal/Fetal Monitor A-3
2036946-001
Factory Defaults: Table of Defaults
Table of Defaults
Summary of Factory Defaults
Setup
Screen
Field Description Factory Default Default Options Hospital Preference
FECG or
US/US2
FHR Volume 5 0-9
FHR Alarm Limits High Low
160 120 bpm
200-140, Off
60-140, Off
Audio Alarms On On, Off
Volume 5 1-9
NIBP Initial Target Pressure 135 mmHg
(18.0 kPa)
100-250 mmHg (13.3-33.3
kPa) in increments of 5
mmHg (0.7 kPa)
Mode Manual Manual, 1, 2, 3, 4, 5, 10, 15,
20, 30, 40, 45, 60, 90, 120
min
NIBP Done Vol 5 0-9
Alarm (mmHg mode)
Systolic
Diastolic
MAP
MHR/P
High Low
160 90 mmHg
90 50 mmHg
140 50 mmHg
120 50 bpm
High Low
70-240 50-150
70-130 30-120
70-150 30-120
100-250 35-120
Alarm (kPa mode)
Systolic
Diastolic
MAP
MHR/P
High Low
21.3 12.0 kPa
12.0 6.7 kPa
18.7 6.7 kPa
120 50 bpm
High Low
9.3-32.0 6.7-20.0
9.3-17.3 4.0-16.0
9.3-20.0 4.0-16.
100-250 35-120
Volume 5 1-9

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 250cx Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals