2018112-069 Rev. F eBike II, eBike II L, eBike II EL - 127 -
• Disconnect the power connector.
• Disconnect the motor cables.
• Disconnect the remote control cable connector the
same way.
• Loose the xing screws of the motor control unit.
• Assembly of the new control unit is performed in
reverse order.
Hint
• Do not interchange the motor connectors!
(Cable and jacks are marked !)
• Following the exchange of a motor or after any malfunc-
tion the control unit has to be initialised!
To initialise the control unit, both motors have to be
moved in dened end positions::
- The motor for the supine angle has to be retracted
completely (supine horizontal)
-- The motor for the supine angle has to be extended
completely (supine 45° to the side)