2018112-069 Rev. F eBike II, eBike II L, eBike II EL - 73 -
nIBp CalIBratIon
This function allows the calibration of the internal pressure
measurement system.
Select „NIBP Calibration“ in the service menu and conrm
with ENTER (or SELECT).
Oset calibration
During this calibration step NO pressure should be present
in the system.
The cu connector (self closing) has to be opened (use cu
connector).
Remove cu
Connect the cu connector with a short tube
The actual pressure value is displayed.
Press SET (or START)
The zero value is stored
Calibration with 200 mmHg
Connect the complete measurement system (hand pump,
manometer and compensating container) as described
under „NIBP test“.
Using the hand pump, apply a xed pressure of
200 mmHg to the system
(control by the external manometer)
Wait some seconds until the pressure is stable -
correct the pressure if necessary
Press SET (or START)
The NIBP calibration is nished
preSSure calibration - offSet
NIBP Calibration
Offset: 0 mmHg
SET
preSSure calibration - gain
NIBP Calibration
Apply 200 mmHg
SET