2018112-069 Rev. F eBike II, eBike II L, eBike II EL - 51 -
nIBp paraMeter
If the ergometer load is controlled by the internal
programs, this parameter denes the start of the NIBP
module.
Leadtime
Denes the leadtime of the bloodpressure measurement to
the next load step
Setting range: 40 to 90 sec. in 1 sec. steps
Resting NIBP
Denes whether or not a resting bloodpressure
measurement is taken before starting the stress test.
Setting range: OFF - ON
pC-Mode
PC-Mode „Automatic“
If „YES“ is activated, the ergometer automatically switches
to remote mode (PC mode) as soon as a valid signal is
received from the interface.
saddle CalIBratIon
(eBike II comfort only)
This parameter is used to adjust the electrical saddle
motor. (display scale 1-40)
Select „TOP Position“ and conrm with „ENTER“.
By use of the arrow keys, move the saddle to the desired
top position (standard 36 cm above saddle guidance) and
conrm with „SET“.
Select „Bottom Position“ and conrm with „ENTER“.
By use of the arrow keys, move the saddle to the desired
bottom position (standard 7 cm above saddle guidance)
and conrm with „SET“.
configuration of nibp paraMeterS
NIBP Parameter
Leadtime
60 sec
Resting NIBP
ON
ENTER
autoMatic Mode
PC Mode
Automatic
Yes
No
ENTER
calibration of electrical Saddle adJuStMent
Saddle Calibration
TOP Position
BOTTOM Position
ENTER
Saddle Calibration
TOP Position
BOTTOM Position
SET