2018112-069 Rev. F eBike II, eBike II L, eBike II EL - 49 -
The service menu with the following parameters appears:
Baudrate 8N2
Static Load Calibration
Analog Comparison (with option COM only)
ECG-Startpulse
NIBP Parameter (with option NIBP only)
Software Update
Dynamic Load Calibration
PC-Mode
Saddle Calibration (eBike II comfort only)
NIBP Calibration (with option NIBP only)
NIBP Test (with option NIBP only)
Ergometer Type (without option NIBP only)
System Cong.
Error log
Select the desired parameter by use of the right and left
softkeys (↑ ↓) and conrm with SET.
default settIngs
settIngs Menu
default mode menu
protocols see operator manual
contrast at optimum
load change +/- 5W
language US devices english
other devices german
• beep ON
date/time current values
ECG type digital
rpm min: 54 / max: 65
pulse display OFF
servICe Menu
Baudrate 8N2 4800
ECG-Startpulse leadtime: 60 sec
pulswidth: 100 msec
NIBP Parameter leadtime: 60 sec
Resting NIBP: ON
PC-Mode-Automatic: ON
Ergometer Type ER900
Analog comparison 20W = 0,2 V / 300W = 3.0 V
System Cong.
Saddle engine eBike basic/L/EL 0
eBike comfort 1
NIBP module devices w/o NIBP 0
devices with NIBP I
Analog interface 0
60 Hz Power US devices 1
other devices 0
Hint
The service menu in the ergometers is available in english
only.
Depending on the actual ergometer conguration, not all
settings are visible.
Leaving the service menu is only possible by switching the
device o.
The sequence of the following descriptions does not
follow the sequence in the service menu.
Selection of paraMeterS in Service Menu
Service Menu
Baudrate 8N2
Static Load Calib.
Analog Comparison
ECG-Startpulse
NIBP Parameter
SET