2018112-069 Rev. F eBike II, eBike II L, eBike II EL - 49 -
The service menu with the following parameters appears:
• Baudrate 8N2
• Static Load Calibration
• Analog Comparison (with option COM only)
• ECG-Startpulse
• NIBP Parameter (with option NIBP only)
• Software Update
• Dynamic Load Calibration
• PC-Mode
• Saddle Calibration (eBike II comfort only)
• NIBP Calibration (with option NIBP only)
• NIBP Test (with option NIBP only)
• Ergometer Type (without option NIBP only)
• System Cong.
• Error log
Select the desired parameter by use of the right and left
softkeys (↑ ↓) and conrm with SET.
default settIngs
settIngs Menu
• default mode menu
• protocols see operator manual
• contrast at optimum
• load change +/- 5W
• language US devices english
other devices german
• beep ON
• date/time current values
• ECG type digital
• rpm min: 54 / max: 65
• pulse display OFF
servICe Menu
• Baudrate 8N2 4800
• ECG-Startpulse leadtime: 60 sec
pulswidth: 100 msec
• NIBP Parameter leadtime: 60 sec
Resting NIBP: ON
• PC-Mode-Automatic: ON
• Ergometer Type ER900
• Analog comparison 20W = 0,2 V / 300W = 3.0 V
• System Cong.
Saddle engine eBike basic/L/EL 0
eBike comfort 1
NIBP module devices w/o NIBP 0
devices with NIBP I
Analog interface 0
60 Hz Power US devices 1
other devices 0
Hint
• The service menu in the ergometers is available in english
only.
• Depending on the actual ergometer conguration, not all
settings are visible.
• Leaving the service menu is only possible by switching the
device o.
• The sequence of the following descriptions does not
follow the sequence in the service menu.
Selection of paraMeterS in Service Menu
Service Menu
Baudrate 8N2
Static Load Calib.
Analog Comparison
ECG-Startpulse
NIBP Parameter
SET