- 78 - eBike II, eBike II L, eBike II EL 2018112-069 Rev. F
eBIke II BasIC / CoMfort
preventIve MaIntenanCe
fInal CHeCkout proCedure / funCtIonal test
Check the setup for stability, adjust with levelling device if required.
Expected result: No instability Pass/Fail
Visually inspect the device:
- inspect coat of lacquer
- inspect cables / power cord
- Check bellows on steering column and saddle
Expected result: No damage or wear detected Pass/Fail
Check cranks, pedals and Velcro strips
Expected result: Cranks, pedals and velcro strips OK Pass/Fail
eBike basic
- Check handlebar adjustment (angle)
- Check saddle height adjustment
- Check that the clamping levers (saddle (height) and handlebar (angle) adjustment) are tight.
Grease the threads of the clamping levers (every 2 to 6 months, depending on applicationpurpose).
(Use of an universal high-performance lubricant such as OKS 470 is recommended)
eBike comfort
- Check electrical saddle adjustment: adjust minimum and maximum saddle height;
then return saddle to mid-level
- Check correct indication of saddle height display
- Check handlebar adjustment (angle)
- Check that the clamping lever for handlebar adjustment (angle) is tight
Expected result: Adjustments work and levers are tight, all threads properly greased Pass/Fail
Using the power cord, connect the device to the power line and turn it on with the power switch
Wait for selftest to end:
- check LCD (e.g. for error messages)
- check in the software if the rmware version is OK
Expected result: Device powers on without errors and FW is OK Pass/Fail
Hint
eBike II basic / comfort is a protection class II device (no connection to protective ground)
preventive maintenance has to be performed every 2 years