- 66 - eBike II, eBike II L, eBike II EL 2018112-069 Rev. F
Hint
• To perform a static load calibration, a 8 kg calibration
weight is needed.
statIC CalIBratIon wItH CalIBratIon weIgHt
The alignment and calibration procedures (static load
calibration, dynamiv load calibration, NIBP calibration) are
identical for all eBike II ergometers.
All calibration processes are started in the service menu.
To activate the service menu the ergometer has to be
switched o and on again.
As soon as the self test appears on the ergometer display,
press the two softkeys
synchronously and hold:
To prevent settings from being changed by chance, the
service menu is „protected“ by a password.
Select „999“ with the arrow keys and conrm with SET.
The service menu appears.
Select „„Static load calibration“ with the arrow keys and
conrm with ENTER.
Selftest running
Service Menu control terMinal
paSSword for Service Menu
Service Code
999
SET
Service Menu control terMinal
Service Menu
Baudrate 8N1
Static Load Calib.
Analog Comparison
ECG-Startpulse
NIBP Parameter
ENTER