- 136 - eBike II, eBike II L, eBike II EL 2018112-069 Rev. F
error Code BIts Bde (Boodpressure)
Bit Error Reaction Display / Info Troubleshooting
0 not used
1 cuff loose Warning - check correct placement of cuff
at patient (tight)
- check cuff
2 Micro position Warning - check correct placement of
microphone (see User Manual)
- check microphone in cuff
- replace cuff
3 Leakage Warning - check cuff
- check hoses, hose connections
- check compensation container
- replace pneumatic unit
4 Errror Calibration Module is switched off Internal error - perform NIBP test
- perform NIBP calibration
- replace BDE PCB
5 Error offset Module is switched off Internal error - perform NIBP test
- perform NIBP calibration
- replace BDE PCB
6 not used
7 not used
8 Safety state mode Module is switched off Internal error - restart eBike (OFF / ON)
- if error persists:
replace BDE
9 Setup not ready Module is switched off Internal error - restart eBike (OFF / ON)
- if error persists:
check cable ZME <-> BDE
- replace BDE
10 CAN zykl. Timeout Module is switched off Internal error - restart eBike (OFF / ON)
- if error persists:
check cable ZME <-> BDE
- replace BDE