- 92 - eBike II, eBike II L, eBike II EL 2018112-069 Rev. F
eBIke II l
preventIve MaIntenanCe
fInal CHeCkout proCedure / funCtIonal test
Visually inspect the device
- Inspect coat of lacquer
- Inspect cables/power cord
- Inspect bellows on steering column and saddle
- Inspect footboard for stability
- Inspect cu, tubing and tubing connection
- Inspect couch adjustment motor for damage
- Check the setup for stability, lock castor brakes, if required.
Expected result: No damage, instability or wear detected Pass/Fail
Check rests, cranks, pedals and Velcro strips
- Check that all clamping levers and star grips are rmly seated
- Check that cranks, pedals, Velcro strips and pedal shoes are properly xed.
- Inspect leg rests and their attachment, if leg rests are installed.
- Inspect head and arm rests as well as handgrips.
- Grease the threads of the clamping levers (every 6 months, depending on application purpose).
(Use of an universal high-performance lubricant such as OKS 470 is recommended)
Expected result: All threads properly greased, cranks, rests and pedals OK Pass/Fail
Power on
- Using the power cord, connect the device to the power line and
turn it on with the power switch.
- Wait for self-test to end:
Check LCD, e.g. for error messages.
- Check in the software if the rmware version is OK.
Expected result: Device powers on without errors and FW is OK Pass/Fail
Couch adjustment motor
- Set motor to both stop positions
- Change couch position from horizontal (0°) to semi-recumbent (45°)
and then back to thehorizontal position (0°)
Expected result: Couch position changes to the expected positions Pass/Fail
Hint
eBike II L is a protection class I device (with connection to protective ground)
Suitable common ground terminal:
screw head of the protective earth connection between swivelling section and ergometer base
preventive maintenance has to be performed every 2 years