- 210 - eBike II, eBike II L, eBike II EL 2018112-069 Rev. F
teCHnICal speCIfICatIons eBIke II Blood pressure Module
Measuring Method: auscultatory method, oscillometric; for resting BP, the
results from both measurements are compared for
plausibility
Measuring Range systolic pressure: 40 to 280 mmHg
diastolic pressure: 40 to 280 mmHg
pulse rate: 35 to 230 P/min
Measurement Error pressure readout error: +/- 3 mmHg
readout resolution: +/- 1 mmHg
Ination Pressure 300 mmHg max.; during ination the ination pressure
automatically adapts to patient's BP
Ination Rate between approx. 6 seconds (to 140 mmHg) and
approx. 18 seconds (to 300 mmHg)
Max. Cu Pressure 300 mmHg
Cu Deation Method pulse-dependent deation rate
approx. 3 mmHg/beat or approx. 3 mmHg/s
Calibration calibration with external pressure meter
Artifact Rejection automatic artifact rejection and comparison of the
resting BP readings from both methods for plausibility