GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
4-16 Section 4-2 - General Procedure
4-2-7 Archiving and Loading Presets for BT06/V65x or lower
NOTE: Always save presets before any software reload. This ensures the presets loaded after the software
reload are as up–to–date as possible.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an CD-
R/MO/DVD-R disk for reloading on the system.
4-2-7-1 Formatting CD-R/MO/DVD-R Disk
1.) Insert an empty (blank) CD-R/MO/DVD-R disk into the CD-R/MO/DVD-R device.
2.) Access to the Utility Menu on the Touch Panel, and select Connectivity>Removable Media or
Tools.
NOTICE
Presets should NOT be saved on the same CD-R/MO/DVD-R disk as images. The Archive Menu
lists the images but does NOT list the presets stored on a CD-R/MO/DVD-R disk.
Figure 4-19 Formatting Removable Media1