DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
8-28 Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-8-6 Date and Time Setting (cont’d)
5.) Set the current date and time.
6.) Click on Apply.
7.) Click on OK to close the Date and Time window.
8.) Select Start > Shutdown > OK to power OFF the scanner.
Figure 8-86 Date & Time
Month and Year
Day
Time