DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
8-32 Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-10 System Settings
After application is successfully installed, various LOGIQ™ 7 system parameters must be properly set.
Table 8-7 System Settings
Item Method Proper Value
Software Version Utility
\
System
\ About
s/w version : R8.x.x
Image Part No. :
5305206-3
Regional Preset Utility
\ System
\
About
\ Additional About Information
For the system other than “NONE”
selected as Regional Preset, resetting
MUST be required.
Computer Name Utility
\
Connectivity
\ TCP/IP
\ Computer Name
\
Enable DHCP
\ IP Address
\
Subnet Mask
\ Default Gateway
Same as Serial Number
OFF
restore customer setting
restore customer setting
restore customer setting
Patient
Database Check
Utility
\ Connectivity
\
DataFlow
\ Default Data Flow
Press Patient (Keyboard or Touch Panel)
Local Archive – Int. HD
(n/a – Automatic)
LCD Monitor Setting Press ‘Toggle Button’ for few seconds (various)
Software Option Utility
\ Admin
\ System Admin
\
HW Number Same as Option Dongle ID
VCR Utility
\
Peripherals
\
Video Setting
\
VCR
PAL/NTSC-depends on region
(set accordingly)
Language pre-work (For
Japanese Only)
Utility
\
Regional Option
\
Language (tab)
\
Install Files to East Asian Language
(Reboot)
ON / Apply