DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-17
8-2-8 Modify System/Windows Settings
After Base System is successfully installed, various parameters must be properly set before proceeding to Application
load. The below outlines necessary system/windows setting.
8-2-8-1 UPS Setting
UPS automatically setup by Ghost. You do not need to setup it manually. You can see the message
below on the desktop.
Please choose OK.
Table 8-6 System/Windows Settings
Item Method Proper Value
UPS Setting Automatic Setting (n/a – just load)
(Restart System)
Avance AC97 Audio Load from c:\windows\system32\drivers (n/a – just load)
(Restart System)
Drive Letter Automatic Setting but check desktop icon if not. DVD Drive = G
HW Device My Computer
\
Properties
Various
LAN Desktop
\
My Network Places
\ Properties
\
Local Area Connection Double Click to activate
TCP/IP Setting Desktop
\
My Network Places
\ Properties
\
Local Area Connection
\ Internet Protocol (TCP/IP)
\
Properties
\ Advanced
\ Options
\
TCP/IP Filter Values in ‘Permit Only’
Date/Time Start
\
Settings
\ Control Panel
\
Date and Time
Proper Time Zone
Accurate Date / Time
Figure 8-69 UPS Setting