DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
8-10 Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-6-4 Saving Connectivity
Before starting upgrading, write down
the Connectivity Setting for back-up.
1.) Touch Utility > Connectivity on the touch panel and click the TCP/IP tab.
2.) Write down the following parameters:
Figure 8-60 TCP/IP
Parameters Descriptions (Entry for Stand-Alone System)
Computer Name
Enable DHCP This parameter
must NOT be selected.
IP-Address
(192.168.1.2)
Subnet Mask (255.255.255.0)
Default Gateway (192.168.1.1)
Note : For a stand-alone system, preset values of IP adress, subnet mask, and default gateway, shown above,
must entered.
Table 8-3 TCP/IP
Remove the
Check mark.