EasyManua.ls Logo

GE LOGIQ 7

GE LOGIQ 7
394 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 4-4 - Application Turnover Check List 4-27
Section 4-4
Application Turnover Check List
Complete these checks before returning the scanner to customer for use:
4-4-1 Software Configuration Checks
Table 4-9 Software Configuration Checks
Step Task to do Expected Result(s)
1.
Check Date and Time setting Date and Time are correct
2.
Check that Location (Hospital Name and
Department) is correct
Location Name is correct
3.
Check Language setting Language is proper
4.
Check Units setting Units are proper

Table of Contents

Other manuals for GE LOGIQ 7

Related product manuals