EasyManua.ls Logo

GE LOGIQ 7

GE LOGIQ 7
394 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
5-42 Section 5-7 - Air Flow Control
Section 5-7
Air Flow Control
5-7-1 Air Flow Distribution
The four air flow pathes allow the scanner to be cooled down as shown in the figure above.
Path A (Front lower left > Filter > PC Box > Rear upper left) for PC Box cooling.
Path B (Front lower left > Filter > LV unit > Rear lower left) for LV unit cooling.
Path C (Bottom right > Filter > NEST Assy > Rear upper right) for NEST Assy cooling.
Path D (Rear bottom right > Filter > HV unit > Rear upper right) for HV unit cooling.
5-7-2 Filters
The scanner contains the three filters located at:
- Front lower left for air flow of the LV unit and PC box.
- Bottom right for air flow of the NEST Assy.
- Rear bottom right for air flow of the HV unit.
Figure 5-67 Air Flow Inside the Scanner
A
B
C
D
A
B
C
D

Table of Contents

Other manuals for GE LOGIQ 7

Related product manuals