GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
5-28 Section 5-5 - Common Service Platform
5-5-2-2 Service Login (cont’d)
4.) Select GE Service at the “Select User Level” field.
5.) Enter the password for the Service Platform.
6.) Click on Okay.
5-5-2-2-2 For BT03 or lower:
Click on the button with the wrench icon in the status bar at the bottom of the scanner console. This button links
the user or the Field Engineer (FE) to the service login screen.
5-5-2-3 Access / Security
The service interface has different access and security user levels. Users are only granted access to
the tools they are authorized to use.
NOTE: *For a GE Field Engineer, the password changes at specific intervals.
Except for iLinq, all access to the service interface is via the Network port. A modem on the scanner is
specifically intended for iLinq use.
Every access request, whether successful or not, will be logged into a service access log that is
viewable to authorized users.
Figure 5-48 User Level
Table 5-2 User Level
User Level Access Authorization
Operator
These service tools are normally used in-house and are protected with a special scanner configuration
that is not allowed to leave the facility.
Administrator
Use the OnLine Center access method provided by
iLinq.
External Service
Required for a external Service other than GE FE.
GE Service
Requires a network connection and knowledge of a password.*