GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-9
8-2-6 Pre-installation Procedures
The following section describes activities necessary before system and application installation.
8-2-6-1 Disabling USB Devices
1.) Turn OFF all USB peripherals connected to the scanner.
8-2-6-2 Save Customer Data / Setting
Save the following user data.
- Patient image data (Refer to operation manual.)
Please confirm the user if the user has the Media (CD/DVD-R) that is not finalized on the system.
You should finalize the Media before R8.x.x Upgrade.
Because once the system is upgraded to R8.x.x, The system never recognizes the no-finalized
media (CD/DVD-R).
8-2-6-3 Operator Login Password
The Operator Login Password will be deleted even if the BECOMP is not replaced. So, write down the
password if necessary.
CAUTION
Loading R8.x.x application software procedures requires basically C drives to be ghosted. In
this case, patient image data, preset parameters, and customer data will be kept. However , in
case full ghost will be required, you MUST back up before starting this procedure as follows.
For customer data safe, we recommend you to back up even for the C ghost only.
Figure 8-59 Login Password
Parameters Descriptions
Operator Login Password
Table 8-2 Login Password

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals