DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-31
8-2-9-1 Installing R8.x.x Application Software (cont’d)
6.) Type Y key.
NOTE: Press the N key to cancel this procedures and to power OFF the system.
NOTE: When the Information Window appears asking you to select OK, do nothing. This window will
disappear automatically.
7.) The installation procedures will start automatically. It will take 10 minutes.
8.) After the installation procedures are completed, the system will power OFF automatically.
Figure 8-89 Y
NOTICE
Do not operate with using mouse or keyboard during installation
procedures!!! The system shall be restarted automatically during
installation.