DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-15
8-2-7 Install Base System (cont’d)
4.) The following screen appears. Press
B then Enter to initialize the C drive
5.) Check if the message appears as reloading C: partition only, then press Enter key.
If the message appears as reloading the entire disk, press CTRL-ALT-DEL to reboot the sysytem
and retry from procedures described in previous page.
Figure 8-65 Select B
CAUTION
If you select [A], all the archive would be lost.
Figure 8-66 Enter