DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-37
8-2-10-3 Computer Name and Connection Settings (cont’d)
7.) Touch Scan on the Touch Panel to return to the scan screen.
8.) Shut down the scanner.
9.) Power ON the scanner, then press Patient key and verify that the following screen appears.
Figure 8-96 Patient