Page #264 background imageLoading...
Page #264 background image
GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
8-58 Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-11-1 Service Platform (cont’d)
2.) Make sure that CAPS is OFF (should be dimmed)
for password entry performed later.
3.) The Service Login window for Service Platform will be shown on the monitor display.
4.) Select GE Service at the “Select User Level” field.
5.) Enter the password for the Service Platform.
6.) Click on Okay.
Figure 8-123 Service Login
CAUTION
If the Login window for Service Platform is not displayed on the monitor, the installation of the
Service Platform has failed. Reload the application software or Base System Software (OS) +
application software. Contact a Technical Support for details.
Figure 8-124 User Level
OFF
ON