Safety Tips
Chapter 1: Introduction
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 1 - 5
Service Manual (MAN272 REV 4)
1
CPR Release
Only healthcare professionals should use the emergency CPR release. The two
release handles are located under the head section of the bed near the seat
section.
To activate the CPR release, pull the handle away from the bed. Continue to
pull out on the handle until the head section is flat. When this is complete,
attend to the patient. The bed will automatically flatten the head and seat
sections and reset itself to be ready to use after the emergency.
WARNING:
A bed that is transported with the steer caster locked in an incorrect
position can drift from side to side during transport. Personal injury or
equipment damage could occur.
WARNING:
Only facility-authorized maintenance personnel should troubleshoot the
bed. Troubleshooting by unauthorized personnel could result in
personal injury or equipment damage.
WARNING:
Unplug the bed from its power source and disconnect the battery
back-up before checking ohms/resistance measurements. Failure to
disconnect line voltage to the bed can damage the VOM and cause
equipment damage or personal injury.
WARNING:
Refer to your VOM owner’s manual for complete and detailed
information regarding the operation of your VOM. Failure to do so could
result in personal injury or equipment damage.
WARNING:
You must use the 2" x 4" pieces of lumber to support the bed. The bed
will fall during this procedure if not supported. Failure to do so could
result in personal injury or equipment damage.