2.2 Hilow Up Malfunction
Chapter 2: Troubleshooting Procedures
Affinity® Three Birthing Bed and Affinity® Four Birthing Bed Page 2 - 13
Service Manual (MAN272 REV 4)
2
7. Set your VOM to measure V DC. At the logic control P.C. board connector
P15, place your black probe into pin 1 (black wire) and your red probe into
pin 2 (blue wire). Press the hilow up switch. The voltage is between
25V DC and 30V DC.
Yes No
Replace the logic control P.C. board (refer to procedure 4.23).
If this solves the problem, go to “Final Actions” on page 2-10.
Otherwise, continue to step 8.
8. Replace the hilow motor (refer to procedure 4.3).
This solves the problem.
Yes No
Contact Hill-Rom Technical Support.
9. Go to “Final Actions” on page 2-10.