4.3 Hilow Drive Motor Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 10 Affinity® Three Birthing Bed and Affinity® Four Birthing Bed
Service Manual (MAN272 REV 4)
15. Remove the two clevis pins (J) from the hilow drive motor assembly (F),
and from the bed frame.
16. Remove the hilow drive motor assembly (F).
Replacement
1. To install the replacement hilow drive motor assembly (F), reverse the
removal procedure.
2. Do the “Function Checks” on page 2-4.