3.7 Using D/A Output
96
Phase angle
Voltage/ current ripple rate, total harmonic voltage/ current distortion
Voltage/ current crest factor
STD.2 (2 Vf.s.) settings STD.5 (5 Vf.s.) settings
STD.2 (2 Vf.s.) settings STD.5 (5 Vf.s.) settings
Although the voltage ripple rate, current ripple rate, total harmonic voltage distortion, and total harmonic cur-
rent distortion are displayed up to 500.00%, level output changes to +9 V at 450% when using the STD.2
(2 V) setting and 180% when using the STD.5 (5 V) setting, voltages in excess of that value are not output.
STD.2 (2 Vf.s.) settings STD.5 (5 Vf.s.) settings
Although the voltage crest factor and current crest factor are displayed up to 612.00, level output changes to
+9 V at 45 when using the STD.2 (2 V) setting and 18 when using the STD.5 (5 V) setting, voltages in excess
of that value are not output.