Checking Voltage Error (Voltage Monitor Check Function)
8.4 Checking Voltage Error (Voltage Monitor Check
Function)
A warning can be displayed on the screen, if a voltage value measured internally by the instrument
becomes abnormal.
Display with the voltage monitor check function set to on
Resistance value measurement mode
(initial screen)
[V.CHK] is displayed on the measurement screen under
normal conditions.
If the voltage is outside the set range (error condition),
[V.CHK] is highlighted in reverse video. Because the
measurement was performed before the voltage became
stable in this case, increase the delay time or extend the
voltage range for the alarm.
Even when the voltage monitor check is activated, the measurement will be completed normally, and the
data will be output.
Procedure to display the setting screen: (Measurement screen) MENU key > [SYS] tab
1
Select whether to generate an alarm if a voltage error occurs.
[ON] Enabled
[OFF] Disabled (default setting)
2
If [ON] is select, set the voltage range.
±2% to ±20% (default setting: ±20%)
Press the F3 key [↓] or F4 key [↑] to change the value.
8
System Setting