113
External Input/Output Terminals and Signals
Output signal
EOM End of measurement Outputs the On signal when the measurement and the
judgment are completed.
INDEX Measurement reference Outputs the On signal when the measurement circuit
nishes the A/D acquisition.
ERR Measurement error Outpauts the On signal if a measurement error occurs.
The following instances result in measurement error:
• The A/D converter for measurement and voltage measurement has overowed.
• The contact-check measured value or xture-capacitance open correction value
exceeds 99.999 pF.
C_CHECK_GO Contact check result Outputs the On signal when the contact check gives a
pass judgment.
V_CHECK_GO Voltage error check result Outputs the On signal when any voltage error was not
found.
OPEN_GO Open correction result Outputs the On signal when the open correction gives
a pass judgment.
VON Voltage output status Outputs the On signal while outputting a voltage.
HI Comparator judgment of Hi Outputs the On signal when the comparator gives a Hi
judgment.
IN Comparator judgment of IN Outputs the On signal when the comparator gives an
IN judgment.
LO Comparator judgment of LO Outputs the On signal when the comparator gives a Lo
judgment.
PASS Judgment result of PASS Outputs the On signal when the comparator gives
an IN judgment after the measurement is normally
completed.
FAIL Judgment result of FAIL Outputs the On signal when the comparator gives
a judgment other than IN after the measurement is
normally completed.
9
External Control (EXT I/O)