119
Timing Chart
Output signal status on start-up
After turning on the instrument, when the screen changes from the start-up screen to the measurement
screen, the EOM and INDEX signals change to ON.
INDEX
Splash screen
EOM
ON/OFF
Judgment
result
Measurement screen
TRIG
ON
OFF
Turn on the instrument
Status
Self-calibration
OFF
OFF
OFF
ON
ON
Judgment results: HI, IN, LO, PASS, FAIL, ERR
The above chart indicates the operation when the trigger source is set to the EXT.
Flowchart for acquiring a judgment result or a measured value with
the external trigger setting
The owchart indicates the procedure with the external trigger setting from the start of the measurement to
the acquisition of a judgment result or a measured value. The instrument outputs the EOM signal immediately
after the judgment results (HI, IN, LO, PASS, FAIL, ERR) have been determined. If the response of the
controller’s input circuit is relatively slow, a waiting time is necessary from between the detection of the EOM
signal switching to on and the acquisition of the judgment results.
Starting the
measurement
Judgment result
OFF
Completion of
measurement
Starting the
measurement
End of
measurement
reception
Judgment result
acquisition
The instrument Controller
TRIG
Waiting for EOM
(If using level
detection, after
waiting for
0.5 ms.)
EOM
HI, IN,
LO, PASS,
FAIL, ERR
During
measurement
9