4.1 Starting Measurement After Measured Value
Becomes Stable (Delay Function)
For the case of external trigger (EXT), set the time from when a trigger is input to when a
measurement starts. (The delay time elapses after a trigger is input each time.) With the trigger
source setting set to the internal trigger (INT), no delay time elapses after the trigger. Set it on the
Measurement screen.
1
Press the F4 key [MODIFY] to enable the measurement conditions to be changed.
Pressing the ESC key cancels the changes of the measurement conditions.
2
Set the delay time.
0 ms to 9999 ms (Default setting: 0 ms)
Pressing the F4 key [EDIT] enables the numerical value to be changed in increments of
1 ms.
Settle
Cancel
Move to the next
digit (left or right)
Change the
numerical value
(up or down)
The numerical value can also be changed with the numeric keypad. See “Methods for changing
numerical values” (p. 28).