125
Settings for External Input and Output
9.7 Settings for External Input and Output
The following items for the external input and output can be set.
Setting Description Reference
Trigger logic An eective edge can be selected for the TRIG signal. p. 125
Trigger lter This setting can be used to accept the TRIG signal from the EXT
I/O terminal only if the TRIG signal remains in the on state during
the set response time.
p. 126
GO-signal outputting
logic level
This setting can be used to specify the outputting logic level for
the C_CHECK_GO and V_CHECK_GO signals, which are used
with the contact check and voltage monitor on, respectively
p. 127
EOM signal output mode This setting can be used to specify the output method for the EOM
(end of measurement) signal.
p. 128
Trigger logic
An eective edge can be selected for the TRIG signal. The logic of the on and the o edges
changes depending on the NPN/PNP setting.
[ON EDGE] PNP setting: Rising, NPN setting: Falling
[OFF EDGE] PNP setting: Falling, NPN setting: Rising
Procedure to display the setting screen: MENU key > [I/O] tab
[ON] Measurement starts at an on edge (default setting)
[OFF] Measurement starts at an o edge
9
External Control (EXT I/O)