Measuring a Liquid Sample
5.3 Measuring a Liquid Sample
To measure the volume resistance of a liquid sample, insert an electrode with a xed shape in the
liquid sample or use an electrode that can also function as the container for the liquid sample.
When using an electrode for liquid samples
1
2
3
4
5
1
Guard electrode
2
Counter electrode
3
Liquid sample
4
Supporting insulator
5
Main electrode
Example of connection
Setting the instrument
Settings for calculating the resistivity are required.
See “4.13 Setting Resistivity Calculation (Resistivity Measurement Function)” (p. 73).
Press the MODE key to switch the measurement mode to liquid volume resistivity measurement
mode, and then perform measurement. (p. 39)
For the electrode constant, refer to the Instruction Manual of the electrode for liquid samples.