87
Measurement With Use of Shielding Box
5.4 Measurement With Use of Shielding Box
Be sure to read “Before using the shielding box” (p. 13) beforehand.
Measurement of high insulation resistance, which requires to measure current with high-sensitivity,
may be unstable due to ambient noise or inductive current.
Thus, measurement objects to be measured must placed in a shielding box.
Guard
Shielding box
Connected to the INTER LOCK terminal
Example of connection
Model SME-8350 Shielding Box is provided as an option.
Setting the instrument
For the measurement with use of a shielding box, set the measured value display mode according
to the objects to be measured.
See “5.1 Measuring Components or Circuits” (p. 77), “5.3 Measuring a Liquid Sample” (p. 86),
and “5.2 Measuring Flat Sample” (p. 82)
5
Measurement Methods Suitable for Various Objects to Be Measured