EasyManua.ls Logo

Hioki Super megohm SM7110 - Confirming Judgment Results

Hioki Super megohm SM7110
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
92
Conrming Judgment Results
6.2 Conrming Judgment Results
An indicator appears on the measurement screen according to the judgment results.
The respective parameter judgment result and comprehensive judgment result are output to EXT I/O.
When the measurement is normally completed along with an IN judgment, the On signal is
outputted through the PASS terminal.
If the measured value is smaller than the upper limit value but greater than the lower
limit value that are previously set
If the measured value is greater than the upper limit value that is previously set
If the measured value is smaller than the lower limit value that is previously set
Judgment
result
Measurement result
Judgment
result
Output of EXT I/O
HI IN LO ERR PASS FAIL
(Upper limit) < (Measured value) Hi on o󰀨 o󰀨 o󰀨 o󰀨 on
(Lower limit) ≤ (Measured value) ≤ (Upper limit) IN o󰀨 on o󰀨 o󰀨 on o󰀨
(Measured value) < (Lower limit) Lo o󰀨 o󰀨 on o󰀨 o󰀨 on
Current Over Range Hi on o󰀨 o󰀨 o󰀨 o󰀨 on
Measurement error*
No
judgment
o󰀨 o󰀨 o󰀨 on o󰀨 o󰀨
During interruption of measurement
No
judgment
o󰀨 o󰀨 o󰀨 o󰀨 o󰀨 o󰀨
* The following instances result in measurement error:
The A/D converter for measurement and that for voltage measurement have overowed.
The contact-check measured value or xture-capacitance open correction value exceeds 99.999 pF.

Table of Contents

Related product manuals