161
Troubleshooting
Error No. Display Cause Solution
ERR:113
Can not set this speed.
Please change valid
range.
The specied measurement
speed cannot be set with the
present range setting.
Change the range to another, and
then change the measurement
speed.
See “Current measurement
accuracy” (p. 141).
ERR:114
Can not set this range.
Please change valid
speed.
The specied range cannot
be set with the present
measurement speed setting.
Change the measurement speed
to another, and then change the
range.
See “Current measurement
accuracy” (p. 141).
Measurement errors
Error No. Display Cause Solution
None Current Over Range
The measured value exceeds
the current measuring range.
Set the correct range.
None +Over°C
The measured value exceeds
the temperature measuring
range. The measuring range
is from −40.0°C to 80.0°C.
The temperature is too
high to be measured with this
instrument.
None -Over°C
The measured value is
lower than the temperature
measuring range. The
measuring range is from
−40.0°C to 80.0°C.
The temperature is too
low to be measured with this
instrument.
None Over%rh
The humidity value exceeds
the humidity measuring range.
The measuring range is from
0.0%RH to 90.0%RH.
The relative humidity is too
high to be measured with this
instrument.
Other errors
Error No. Display Cause Solution
None --.--°C
The Humidity Sensor is not
connected.
Connect the Humidity Sensor
sensor to the instrument.
None --.-%rh
The Humidity Sensor is not
connected.
Connect the Humidity Sensor to the
instrument.
12
Maintenance and Service