89
6
Judging Measured Value
(Comparator Function)
The comparator function judges whether measured values are within or out of a range dened by
an upper and lower limits set previously.
A judgment is displayed as follows on the screen.
[Hi] (Upper limit) < (Measured value)
[IN] (Lower limit) ≤ (Measured value) ≤ (Upper limit)
[Lo] (Measured value) < (Lower limit)
Upper and lower limit values
The function judges whether the measured value is higher than the upper limit previously set (Hi),
within the range between the lower and upper limits previously set (IN), or lower than the lower limit
preciously set (Lo).
(Example: For upper limit is 3 G
Ω
, lower limit is 2 G
Ω
, measured value is 1.5 G
Ω
)
2 G
Ω
3 G
Ω
Lower limit value Upper limit value
INLo
Judgment
Hi
1.5 G
Ω
Measured
value
(Lo judgment)
Procedure to display the settings screen: COMP key
1
Select whether the measured value is to be judged or not.
[ON] Yes
[OFF] No (default setting)