Settings for External Input and Output
Trigger lter
This setting can be used to accept the TRIG signal from the EXT I/O terminal only if the TRIG
signal remains in the on state during the set response time.
Procedure to display the setting screen: MENU key >
[I/O] tab
1
Select whether to activate the trigger lter or not.
[ON] Activate the trigger lter.
[OFF] Deactivate the trigger lter (default setting).
2
If [ON] is selected, set the response time (time during which the TRIG signal remains
in the on state).
1 ms to 500 ms (default setting: 1 ms)
Pressing the F4 key
[EDIT] enables the
numerical value to be
changed.
Settle
Cancel
Move to the next
digit (left or right)
Change the
numerical value
(up or down)
Numerical value can also be changed with the numeric keypad. See “Methods for changing
numerical values” (p. 28).