EasyManua.ls Logo

Hioki Super megohm SM7110 - Electrical Specifications

Hioki Super megohm SM7110
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
121
Internal Circuit Conguration
PNP setting
PLC, etc.
Output
Common
Do not connect any external power supply to Pin 8.
The instrument
8
ISO_5 V
1
TRIG
2
C_CHECK
10
ERR
11
C_CHECK_GO
9
ISO_COM
27
ISO_COM
PLC, etc.
Input
Common
2 k
Ω
1 k
Ω
10
Ω
Zener voltage 30 V
Internally isolated common
(This is isolated from the protective ground of the instrument.)
PNP
EXT I/O MODE
selector
Internally isolated
power supply
5 V
30 V max.
Share the ISO_COM for the common terminals of the input and the output signals.
Electrical Specications
Input signal Input type Photo-coupler-isolated, non-voltage contact inputs (supporting
current sink and current source outputs)
Input on Residual voltage 1V or less (Input on current 4 mA [value for
reference])
Input o󰀨 Open-circuit (Breaking current: 100 μA or less)
Output signal Output type Photo-coupler-isolated open drain output (non-polar)
Maximum load voltage 30 V DC
Maximum output current 50 mA/channel
Residual voltage 1V or less (Load current: 50 mA) / 0.5 V or less (Load current:
10 mA)
Internally
Isolated power
supply
Output voltage For sink output: +5.0 V ±10%
For source output: -5.0 V ±10%
Maximum output current 100 mA
External power input None
Isolation Floating from the protective ground potential and the
measurement circuit
Insulation rating Voltage to ground 50 V DC, 30 V rms AC, 42.4 V peak AC or
less