Interface Specications
11.4 Interface Specications
The underlined values are the default settings. The set values are retained even after the instrument is turned o.
USB Communication
contents
Remote control, measured value output
Connector Series B receptacle
Electrical
specications
USB2.0 (Full Speed)
Class (mode)
CDC class (COM mode)
HID class (USB keyboard mode)
RS-232C Communication
contents
Remote control, measured value output
Transmission
method
Start-stop synchronization system, full duplex
Transmission
speed
4800 bps/9600 bps/19200 bps/38400 bps/115200 bps
Data bit length 8 bits
Stop bit 1
Parity bit None
Delimiter Sending: CR + LF; Receiving: CR, CR + LF
Handshake X ow: Not provided, Hardware ow: Not provided
Protocol Non-procedure system
Connector D-sub 9 pins, male, mating xed base screw #4-40
GP-IB Communication
contents
Remote control
Device address
0 to 30, default: 1
Delimiter
LF/CR + LF
Other IEEE488.2 compliant
Interface functions
SH1 Provides all source handshake functions.
AH1 Provides all acceptor handshake functions.
T6 Provides a basic talker function.
Provides a serial poll function.
Does not provide talk-only mode.
Provides a function to release the talker using the MLA (My Listen Address).
L4 Provides a basic listener function.
Does not provide a listen-only mode function.
Provides a function to release the listener using the MTA (My Talk Address).
SR1 Provides all service request functions.
RL1 Provides all remote/local functions.
PP0 Does not provide parallel poll functions.
DC1 Provides all device clear functions.
DT1 Provides all device trigger functions.
C0 Does not provide a controller function.