Interface Specications
Remote functions Operation details If communications are performed through RS-232C, USB, or GP-
IB, the function changes the instrument status to remote and
disables any key operation other than the STOP key.
To deactivate the remote control, follow the ways listed below:
• Press the LOCAL key.
• Turn o the instrument and turn it on again.
• Execute the command
:SYSTem:LOCal
through RS-232C or
GP-IB.
• Execute the command
GTL
through GP-IB.
Communication
monitor function
Operation details Displays the sending and receiving status of commands and
queries.
Setting
ON/OFF
Data output function Operation details With the trigger mode set to EXT:
Outputs measured values when the TRIG signal is input
or the ENTER (TRIG) key is pressed.
With the trigger mode is INT:
Automatically outputs measured values every time
measurement is completed.
Setting
ON/OFF (Can be congured using the panel or through
communications)
Memory function Operation details Stores the measured data in the memory.
Collectively sends the saved measured values.
Number of
memories
999 (volatile memories without backup)
Setting
ON/OFF