Settings for External Input and Output
GO-signal outputting logic level
This setting can specify the outputting logic level for the C_CHECK_GO and V_CHECK_GO
signals, which are used with the contact check and voltage monitor on, respectively (No logic level
of the OPEN_GO signal is inverted).
1
Logic level to be outputted with the NORMAL setting
Procedure to display the setting screen: MENU key >
[I/O] tab
[NORMAL] When a pass judgment is given for the contact check or voltage monitor check: Outputs
the On signal.
When a fail judgment is given for the contact check or voltage monitor check: Outputs
the O signal.
2
Logic level to be outputted with the INVERT setting
Procedure to display the setting screen: MENU key >
[I/O] tab
[INVERT] When a pass judgment is given for the contact check or voltage monitor check: Outputs
the O signal.
When a fail judgment is given for the contact check or voltage monitor check: Outputs
the On signal.
When the contact check or voltage monitor is disabled, the instrument always gives fail judgments
and thus outputs the O signal.