114
Timing Chart
9.3 Timing Chart
The levels of each signal represent whether the contacts are in the on or the o󰀨 state.
With the current source (PNP) setting, the voltage level of the corresponding output terminal
becomes high when an output state is on, and low when a state is o󰀨.
If the output signal setting is set to the current sink (NPN), the voltage levels, high and low are
inverted.
From when measurement starts to when judgment result is acquired
(1) Voltage output, measurement with external trigger setting
VON
INDEX
TRIG
EOM
Voltage output
t3
t4
t6
t5
t7
t8
t9
HI, IN, LO,
PASS, FAIL,
ERR
t10
t11
t13
t12
START
STOP
t1
t2
t14
t15
Contact check delay
Contact
check
Delay
Measurement
V_CHECK_GO
C_CHECK_GO
ON
ON
ON
OUTPUT
ON
ON
ON
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON/OFF
ON/OFF ON/OFF
ON/OFF
ON/OFF

Table of Contents