47
Basic Measurement Examples
3.8 Basic Measurement Examples
Measurement of the resistance value of a capacitor is explained as an example here.
Examples of setting contents
Measured value display mode Resistance value measurement mode (initial screen)
Measurement voltage [100 V]
Current range [AUTO]
Measurement speed [SLOW2]
Set the trigger source to the internal trigger (INT).
1
Press the MODE key to display the Resistance measurement screen. (p. 39)
The instrument displays values that include the input resistance of 1k
Ω
(±10%).
2
Press the VOLT key or VOLT key to set the measurement voltage to [100 V].
(p. 41)
3
Press the SPEED key to set the measurement speed to [SLOW2]. (p. 43)
3
Basic Measurement