Setting Applied Measurement Voltage
3.2 Setting Applied Measurement Voltage
If using the internal power supply
Press the VOLT▲ key or the VOLT▼ key on the measurement screen to change the applied
measurement voltage.
Pressing the F4 key [MODIFY] enables also the function keys to be used to select the voltage.
IMPORTANT
If the measurement voltage setting or the charge setting is changed after the current limiter has
been set and if the current limiter exceeds the upper limit, the current limiter is set to the largest
value among the settable values.
See “4.9 Limiting Current Flowing Through Measured Object (Current Limiter)” (p. 65).
The numerical value can also be changed with the numeric keypad.
See “Methods for changing numerical values” (p. 28).
1
Select the voltage.
0.1, 0.5, 1, 2.5, 5, 10, 25, 50, 100, 250, 500, 1000, 1500*, 2000* (Default setting: 0.1)
* Can be set only for Model SM7120
Pressing the F4 key [EDIT] enables the numerical value to be changed in increments of 0.1 V.
Settle
Cancel
Move to the next
digit (left or right)
Change the
numerical value
(up or down)