Setting Voltage Value for Resistance Calculation
4.4 Setting Voltage Value for Resistance Calculation
Although resistance values are calculated from currents and voltages measured by this instrument,
you can also set the voltage to any optional value and enables the instrument to calculate
resistance values.
Procedure to display the settings screen: (Measurement screen) MENU key >
[MEAS] tab
1
Select the voltage to be used for calculating resistance values.
[EXT.V] Use any optional voltage. (when the external power supply is used)
[VMONI] Use the voltage measured by the instrument (voltage monitor value).
[MES.V] Use the voltage set as the applied voltage for measurement (default setting).
If high-speed measurements are carried out with the V mode set to [VMONI], resistance values can be
measured to be lower than actual values because measurements start before measurement voltages
suciently increase.
2
If [EXT.V] is selected, set the voltage value.
0.1 V to 5000.0 V (default setting: 0.1 V)
Pressing the F4 key
[EDIT] enables the
numerical value to be
changed.