55
Automatic Inspection (Sequence Program Function)
In the following case, the actual measurement time the sequencing measurement requires
becomes longer than the measurement time specied in the [MEAS] eld.
When the specied time of the sequencing measurement is shorter than the interval calculated
from the measurement speed setting specied in the [SPEED] eld
Example:
[SPEED]=FAST(4.1 ms), [MEAS]=1 ms
t
t
4.1 ms 4.1 ms 4.1 ms 4.1 ms 4.1 ms 4.1 ms
24 ms
4.1 ms
1 ms
Actual measurement
time
Practically, the measurement lasts 3.1 ms longer than the specied time.
Sequencing
measureme
nt
[MEAS] setting
t
As described above, the overall time of the sequencing measurement becomes longer, which does
not aect charging and discharging times.
7
Set the discharge time after measurement completes.
Press F3 key [ms/s] to switch the unit.
0 ms to 999.9 s (default setting: 0.0 s)
Pressing the F4 key
[EDIT] enables the
numerical value to be
changed.
During discharge states (and also stopped states), the circuit between the measurement input terminal
and the OUTPUT terminal is connected to another circuit consisting of the current limiter and the input
resistance (approximately 1 k
Ω
) in the current measurement block that are connected to each other in
series.
4
Applied Measurement