Automatic Inspection (Sequence Program Function)
5
Set the charging (voltage application) time before a measurement starts.
Press F3 key [ms/s] to switch the unit.
0 ms to 999.9 s (default setting: 0.0 s)
Pressing the F4 key
[EDIT] enables the
numerical value to be
changed.
6
Set the measurement time.
Press F3 key [ms/s] to switch the unit.
1 ms to 999.9 s (default setting: 0.1 s)
Pressing the F4 key
[EDIT] enables the
numerical value to be
changed.
The time specied for sequencing measurement
For the sequencing measurement, the instrument measures a measuring object repeatedly during
the measurement time specied in the [MEAS] eld at a measurement speed specied in the
[SPEED] eld.
Example:
[SPEED]=FAST(4.1 ms), [MEAS]=24 ms
t
t
4.1 ms 4.1 ms 4.1 ms 4.1 ms 4.1 ms 4.1 ms
24 ms
4.1 ms
1 ms
Actual measurement time
Practically, the measurement lasts 0.6 ms longer than the specied time.
Sequencing
measureme
nt
[MEAS] setting